Abstract
Summary Technologic advances in bronchoscopy continue to improve our ability to perform minimally invasive, accurate evaluations of the tracheobronchial tree and to perform an ever-increasing array of diagnostic, therapeutic, and palliative interventions. The role of both “old” and “new” diagnostic bronchoscopy will continue to evolve as further improvements are made in bronchoscopes, accessory equipment, and imaging technologies. The major challenge in the adoption of the many new bronchoscopic techniques into routine clinical practice is the need for well-designed studies to delineate the appropriate use of these interventions and to better define their limitations.
Introduction
The first bronchoscopy was performed in 1887 by Gustav Killian of Freiburg, Germany [1].
During the early years of the development of bronchoscopy, the indications for the procedure were primarily therapeutic: removal of foreign bodies and dilation of strictures from tuberculosis and diphtheria. In the early part of the 20th century, Chevalier Jackson, the father of American Broncho-Esophagology Association, further advanced bronchoscopic techniques and designed modern rigid bronchoscopies [2]. Again, the primary indication was often therapeutic.
Fibreoptic bronchoscopy (FOB) was developed in the late 1960s by S. Ikeda [3] and has become the mainstay investigation in the evaluation of patients suspected of lung cancer. It is employed mainly as a diagnostic tool providing tissue to determine the histological type of tumour. Bronchoscopy also has a role in disease staging and an extended role in delivering therapeutic modalities. FOB is convenient to perform, safe and well tolerated by the patient.
The requirement of minimal sedation makes it acceptable as an outpatient procedure, thus it has almost completely replaced rigid bronchoscopy in the initial assessment. The development of video bronchoscopes has the added advantage of facilitating teaching and rendering the procedure more interesting for other observers in the bronchoscopy suite.
The flexibility of the bronchoscope allows the operator to inspect the majority of fourth-order and often up to sixth-order bronchi. In addition, the operator may directly assess mucosal details, such as colour and vascularity. Relative contraindications to the procedure are few and include: hypoxaemia refractory to supplemental oxygen (O2), intractable bleeding diathesis, severe pulmonary hypertension, cardiovascular instability and acute hypercapnia [4].
FOB is safe with a complication rate of 0.12% and a mortality rate of 0.04% [5]. The dangers of haemorrhage and pneumothorax relate to the biopsy procedure used and will be discussed later. In all patients, the bronchoscope causes a temporary increase in airflow obstruction, which may result in hypercapnia [6]. Inappropriate sedation with benzodiazepines or opiates will increase the likelihood of respiratory complications and high-risk patients should be identified by prior measurement of arterial blood gases [5–7]. Supplemental O2 should be provided and patients should be monitored throughout with pulse oximetry. Cardiac monitoring should be used for those patients with a history of ischaemic heart disease and resuscitation equipment immediately available.
Although FOB has largely replaced rigid bronchoscopy in the initial assessment of the patient, the rigid scope has advantages in certain situations [8]. As mentioned, it may provide more accurate information regarding the endobronchial location of a tumour prior to resection.
Additionally, manipulation of the scope allows assessment of the mobility of the proximal airways providing an indirect evaluation of mediastinal nodal involvement. Airway obstruction is less and thus the rigid scope may be preferable in exploring patients with tracheal narrowing in whom the flexible scope may produce critical airway narrowing. It provides superior suction, facilitating the assessment and biopsy of potentially haemorrhagic lesions and the debulking of large tumours [8–10]. In addition, many physicians are now relearning the technique to facilitate endobronchial laser therapy and stenting procedures [11].
The diagnostic yield of fibreoptic bronchoscopy
The expected diagnostic yield from FOB depends on the location and distribution of the tumour. Central endobronchial lesions yield the highest diagnostic return (>90%), whilst small peripheral lesions often prove more elusive unless more demanding and time-consuming techniques are used. The question of which combination of cytological and histological procedures provides the optimum diagnostic yield has not been conclusively answered but probably depends on the expertise available in any individual centre. The routine techniques include bronchial washings, brushings and biopsies but these may be augmented by the use of transbronchial needle aspiration (TBNA) and bronchoalveolar lavage (BAL) [12].
More than 70% of lung carcinomas are visible to the FOB and although the yield is dependent on operator experience, a high level of diagnostic accuracy can be achieved by taking between three and five biopsy specimens and a combination of brushing, biopsy and bronchial washes can expect to establish a diagnosis in >60% of cases [6, 7, 13, 14]. When the tumour is visible but is intramural rather than endobronchial in distribution, the diagnostic yield falls to 55% and is reduced further when the tumour lies beyond the bronchoscopist's vision [6, 7, 12].
The main role of BAL in patients with lung cancer is the diagnosis of opportunistic infections in patients undergoing chemotherapy. However, BAL may have an extended role in the diagnosis of malignancy itself. A high diagnostic yield has been shown in the detection of pulmonary haematological malignancies, primary bronchoalveolar cell carcinoma and metastatic adenocarcinoma of the breast [15–17].
Information on the role of BAL in the diagnosis of primary lung cancer remains sparse.
Examination of BAL from 55 patients with a peripheral lung lesion demonstrated a diagnostic yield of ∼30% with no false-positive results and only one instance of incorrect cell typing. Additionally, in combination with bronchial washing and post-bronchoscopy sputum, BAL increased the yield to 56% [18]. Examination of BAL in 162 patients with malignant lung infiltrates revealed improved sensitivity in cases of bronchoalveolar cell carcinoma (93%) and lymphangitis carcinomatosa (83%). 45% of non-Hodgkin's lymphoma was detected and immunocytochemistry was of value in identification and classification [19].
BAL is safe; bleeding and pneumothorax are uncommon and the fever and transient loss of lung function reported are rarely serious and there is no requirement for fluoroscopy. Furthermore, the diagnostic yield is high in diseases other than cancer, such as pulmonary tuberculosis. Advances in cell and molecular biology may complement the technique of BAL to improve the rate of tumour diagnosis in peripheral lesions (particularly adenocarcinoma) and may also provide a useful tool to explore the molecular mechanisms governing the genesis of lung cancer [20–22].
Visible endobronchial lesions
Central tumours can present as exophytic mass lesions, with partial or total occlusion of the bronchial lumen, as peribronchial tumours with extrinsic compression of the airway, or with submucosal infiltration of tumour. The changes with peribronchial tumours or with submucosal infiltration are often subtly, the airways should be examined closely for characteristic changes, such as erythema, loss of bronchial markings and nodularity of the mucosal surface. Central lesions are usually sampled with a combination of bronchial washes, bronchial brushings and endobronchial biopsies. The yield of endobronchial biopsies is highest for exophytic lesions, with a diagnostic yield of ∼90% [23–25]. Three to four biopsies are likely adequate in this situation. Attempts should be made to obtain the biopsies from areas of the lesion that seem viable.
For submucosal lesions, TBNA can be performed by inserting the needle into the submucosal plane at oblique angle, and in patients with peribronchial disease and extrinsic compression, the needle should be passed through the bronchial wall into the lesion [26, 27]. It is particularly frustrating when apparently adequate biopsy specimens from visible endobronchial disease fail to achieve a diagnosis. Reasons for this include the presence of surface necrosis or the presence of crush artefact (particularly common with samples from small cell carcinoma). In these circumstances, TBNA may improve diagnostic yield [28, 29].
Regarding the T staging, FOB may allow the operator to determine that the tumour is beyond resection. Pointers to inoperability include: paralysis of a vocal cord; tumour to the level of the right tracheobronchial junction or to within 2 cm of the left tracheobronchial junction; and definite carinal or tracheal involvement (fig. 1).
Fibreoptic bronchoscopic image showing a tumour that is inoperable due to location.
Peripheral lung lesions
Peripheral lesions are usually sampled with a combination of bronchial wash, brushes, transbronchial biopsy and TBNA. The diagnostic yield of bronchoscopy for peripheral lesions depends on a number of factors, including lesion size, the distance of the lesion from the hilum and on the relationship between the lesion and bronchus. The yield of bronchoscopy for lesions <3 cm varies from 14–50% compared with a diagnostic yield of 46–80% when the lesion is >3 cm [30–32]. The presence of a bronchus sign on chest CT predicts a much higher yield of bronchoscopy for peripheral lung lesions. In these cases, fluoroscopic guidance should be used to ensure proper positioning of the diagnostic accessory (fig. 2).
Computed tomography image of a chest showing a bronchus sign.
Fluoroscopy increases the diagnostic yield from transbronchial biopsy (TBBX) in focal lung lesions but it is time consuming, requires experience and is not universally available. However, if the disease process is diffuse, such as in lymphangitis carcinomatosa, yields are similar whether fluoroscopy is used or not [33]. Indeed, Tbbx may be regarded as the procedure of choice for Iymphangitis carcinomatosa. Complications from TBBX include pneumothorax and haemorrhage but these are generally low and rarely serious.
In situations where bronchial biopsies cannot be obtained, examination of bronchial washings may still yield useful information and often provide complementary information [34, 35]. It is often prudent to perform all types of sampling procedure to maximize the yield [12].
Several studies have demonstrated that TBNA may be used to obtain diagnostic tissue from peripheral lesions. Typical results report an increase in the diagnostic yield from percentage figures in the mid-30s up to the high-60s [12, 36–38]. As with Tbbx, the size of the peripheral lesion appears to be important; although, this is not a feature of all studies. Optimum yields are provided by employing a combination of diagnostic techniques. TBNA may represent an alternative to TBBX when the airway is stenosed or externally compressed to such a degree that it is not possible to pass the TBBX forceps.
The need to workup and manage pulmonary nodules and masses is encountered with increasing frequency in chest medicine. In patients with such nodules, the diagnostic procedure is usually performed as a TBBX under fluoroscopic guidance. The commonly performed procedure, endoscopic transbronchial biopsy, is associated with a low yield in coin lesions <3 cm in size or fluoroscopically invisible lesion problems. The yield depends greatly on the size and location of the abnormality [39, 40] and generally on the ability to visualise the lesion fluoroscopically. Nodules that are too small to be visualised by conventional fluoroscopy during the procedure pose a particular problem and usually require further, often surgical, biopsy procedures.
If issues of diagnostic yield could be resolved, flexible bronchoscopy would represent a very interesting part in the workup and management of lung lesions. What is needed, therefore, are new methods for navigation and localisation, independent of visualisation by fluoroscopy, but still dependent on the technical skills of the bronchoscopist. Promising new technologies are electromagnetic navigation and endobronchial ultrasound.
Electromagnetic navigation
The electromagnetic navigation system is a localisation device that assists in placing endobronchial accessories (e.g., forceps, brush and needle) in the desired areas of the lung.
The system uses low-frequency electromagnetic waves, which are emitted from an electromagnetic board placed under the bronchoscopy table mattress. A 1-mm diameter, 8-mm-long sensor probe mounted on the tip of a flexible metal cable constitutes the main assembly of the device (locatable guide). Once the probe is placed within the electromagnetic field, its position in the X, Y and Z planes, as well as its orientation (roll, pitch and yaw movements), are captured by the electromagnetic navigation system. This information is then displayed on a monitor in real time. The locatable guide also has an added feature that allows its distal section to be steered 360°. The fully retractable probe is incorporated into a flexible catheter (serving as an extended working channel), which, once placed in the desired location, creates an easy access for bronchoscopic accessories. The computer software and monitor allow the bronchoscopist to view the reconstructed three-dimensional computer tomography (CT) scans of the object’s anatomy in coronal, sagittal and axial views together with superimposed graphic information depicting the position of the sensor probe.
There are still some major limitations to the technique. For planning, a CT scan is necessary with a special protocol (1-mm cuts and tight overlay). For the planning of the procedure, the use of the electromagnetic navigation bronchoscopy (ENB) software is required. The planning can be done even on the system or on a special dedicated laptop before the procedure; the planning needs some time, up to 10 min in trained colleagues. The whole procedure time is prolonged compared with a traditional diagnostic bronchoscopy with fluoroscopy; but compared with the technique we have at the moment for lesions of this size, CT-guided punctures, the time is equivalent. Last but not least, the locatable guide is single use and costs US$500–1,000, depending on the market. The reimbursements are also dependent on the local systems and vary US$0–1,000. Schwarz et al. [41] performed a trial to determine the practicality, accuracy and safety of real-time electromagnetic navigation in locating artificially created peripheral lung lesions in a swine model. No adverse effects, such as pneumothorax or internal bleeding, were encountered in any animal in this study. They concluded that real-time electromagnetic positioning technology, coupled with previously acquired CT scans, is an accurate technology that can augment standard bronchoscopy to assist in reaching peripheral lung lesions and performing biopsies.
Based on the results of that study, Becker et al. [42] performed a pilot study in humans. They examined the utility of the system in 30 consecutive patients presenting for endoscopic evaluation of lung nodules and masses. The lesion size in this population varied 12–106 mm but was specifically not controlled for in this early trial. Evaluation was possible in 29 patients and, in 20 patients, a definitive diagnosis was established, with no complications related to the navigation device.
In an uncontrolled study, again Schwarz et al. [43] confirmed that the procedure was safe and added only an average of 15 min to the time of a conventional bronchoscopy. Successful diagnostic biopsies were obtained in 69% of patients. A follow-up study of 60 patients [44], published in 2006, successfully reached the target lesion in 100% of cases. Bronchoscopy with electromagnetic navigation diagnosed 80.3% of the lesions, 74% of the peripheral lesions and 100% of the lymph nodes. Of the lesions, 57% were <2 cm in size. Diagnostic yield did not differ significantly based on the size of the lesion. The accuracy of ENB navigation has been proven in animal studies and against fluoroscopically verified reference points in humans [43, 44]. Nevertheless, all preceding diagnostic studies utilising ENB also used fluoroscopy to guide biopsies. The role of ENB as a standalone technology is still unproven and concerns remain that biopsy instruments may dislodge an accurately positioned extended working channel (EWC) when replacing the sensor probe.
Eberhardt et al. [45] examined the yield of ENB without fluoroscopy in the diagnosis of peripheral lung lesions and solitary pulmonary nodules. 92 peripheral lung lesions were biopsied in the 89 subjects. The diagnostic yield of ENB was 67%, which was independent of lesion size. The mean navigation error was 9±6 mm (range 1–31 mm). When analysed by lobar distribution, there was a trend towards a higher ENB yield in diagnosing lesions in the right middle lobe (88%). The study concluded that ENB could be used as a standalone bronchoscopic technique without compromising diagnostic yield or increasing pneumothorax risk. This may result in sizable time saving and avoids radiation exposure.
Makris et al. [46] confirmed these results. In 40 patients all target lesions but one were reached and the overall diagnostic yield was 62.5% (25 out of 40). Also, the French group summarised that electromagnetic navigation-guided bronchoscopy has the potential to improve the diagnostic yield of transbronchial biopsies without further fluoroscopic guidance and may be useful in early diagnosis of lung cancer, particularly in non-operable patients.
Endobronchial ultrasound
At the moment, two different systems are available. The linear endobronchial ultrasound (EBUS) bronchoscope, which incorporates the ultrasound transducer at its distal end, utilises a fixed array of transducers aligned in a curvilinear pattern.
Owing to the size of the scope, the system is usable only in the central airways. For the peripheral lung the radial EBUS must be used. The radial EBUS system consists of a mechanical radial miniprobe.
Two types of miniprobe are available, one with a notch at the tip for a water-fillable balloon catheter, one without a notch. Particularly with the notch type, ultrasonography can be performed using the balloon method, that is, the balloon is inflated at the distal end of the probe after the probe has been inserted into the instrument channel.
The balloon method makes easy delineation of ultrasound images possible, even at sites where it is hard to retain defecated water. The limitation is the size: with the balloon sheath, an endoscope ≥2.8 mm diameter channel has to be used. The notchless probe is smaller (1.7 mm) and can also be used in smaller scopes.
The 20-MHz frequency is commonly used, although 12- and 30-MHz probes are also available. For use in the peripheral lung, most commonly the probe is placed through a guide sheath in the working channel of the bronchoscope. After localisation of the lesion (fig. 3), the bronchoscope will be kept in place at the nearest visible subsegmental carina and the miniprobe removed. Through the guide sheath, the forceps are guided to the lesion. By using the guide sheath, EBUS-guided TBBX can be performed without losing the position of the nodule. The first studies were done without the guide sheath; nowadays the guide sheath technique is accepted as the technique of choice.
Endobronchial ultrasound showing a lesion.
The feasibility trial of EBUS-guided TBBX [47] without the use of fluoroscopy showed that EBUS can provide an alternative to fluoroscopy for image guidance in biopsies for peripheral lesions. In the study, a trend towards superior results with EBUS was particularly strong in lesions <3 cm in diameter. The same results were shown by Shirakawa et al. [48]. After the feasibility trial, investigators began to examine the use of EBUS as an adjunct for the diagnosis of peripheral lung lesion and solitary pulmonary nodules.
A large prospective study by Paone et al. [49] compared traditional bronchoscopic transbronchial biopsy with EBUS for peripheral lesions. They found that EBUS-guided bronchoscopy had a sensitivity of 0.83 for lesions >3 cm in size and 0.75 for lesions <3 cm in size; compared with traditional transbronchial biopsy, EBUS has also shown promise when used for nodules <3 cm in size. These types of lesion are often difficult to visualise fluoroscopically for transbronchial biopsies, and bronchoscopy has a low diagnostic yield in such settings.
Most of the published studies have used radiographic fluoroscopy, with radiation exposure for both the patient and medical staff. This result is in line with the study by Herth et al. [50], including only peripheral pulmonary lesions <30 mm and reporting a diagnostic yield of 87% for EBUS-guided TBBX without the need for radiographic equipment or radiation exposure.
For lesions <20 mm, the yield of EBUS-guided detection and pathologic diagnosis decreased fewer than 30% yield [51]. By contrast, Japanese groups have reported diagnostic yields of 53 and 72%, respectively, for lesions <20 mm using EBUS-guided TBBX with a catheter sheath and under radiographic fluoroscopy [52, 53]. More recently, one of these Japanese groups have performed EBUS-guided TBBX using virtual bronchoscopic navigation and detected 67% of lesions <20 mm on EBUS, resulting in a diagnostic yield of 44% [54].
The limitation to these systems is a significant learning curve, and methods of physician training and education still need to be worked out. The technique in the peripheral lung is not as complicated as in the central airways, but the whole procedure must be trained. Also, for EBUS, the reimbursement situation has to be discussed with the local system. However, EBUS lacks a navigation system and requires the operator to manoeuvre the bronchoscope blindly to the lesion with the knowledge of prior radiological investigations, such as CT scans.
Biopsies using ENB have not always resulted in a diagnosis despite accurate navigation in the vast majority of cases to within 10 mm of the target centre. Respiratory variations causing larger than anticipated navigation errors and dislodgement of the EWC when biopsy instruments were introduced may account for this lower-than-expected diagnostic yield. ENB lacks a means to directly visualise lesions before biopsy. The role of combining EBUS with ENB to gain the benefits and minimise the limitations of either technique has never been reported. Eberhardt et al. [55] performed a prospective randomised controlled trial comprising three arms with EBUS only, ENB only and combined EBUS/ENB to test this hypothesis.
Of the 120 patients recruited, 118 had a definitive histological diagnosis and were included in the final analysis. The diagnostic yield of the combined procedure (88%) was greater than either endobronchial ultrasound (69%) or electromagnetic navigation alone (59%; p = 0.02).
The group concluded that combined endobronchial ultrasound and electromagnetic navigation improves the diagnostic yield of flexible bronchoscopy in peripheral lung lesions without compromising safety.
Mediastinal staging
Transbronchial needle aspiration
The first description of sampling mediastinal lymph nodes through the tracheal carina using a rigid bronchoscope was by Schieppati [56, 57], an Argentinian physician who presented the technique at the Argentine Meeting of Bronchoesophagology in 1949. In 1978, Wang et al. [58] demonstrated that with this technique it was also possible to sample paratracheal nodes. In 1979, Oho et al. [59] introduced a flexible needle that could be utilised through a bronchofibrescope and in 1983, Wang et al. [60] and Wang and Terry [61] pointed out the diagnostic possibility of the method in staging of lung cancer and developed new types of needles. Subsequent publications highlighted its utility in the diagnosis of endobronchial and peripheral lesions and the ability of TBNA to provide a diagnosis even in the absence of endobronchial disease, in a nonsurgical fashion, confirmed its usefulness to bronchoscopists [62–64].
Operators have reported the use of 21- and 22-gauge cytology needles and 19-gauge histology needles [65]. Further samples are provided by rinsing the needle with a small volume of normal saline and collecting the “flush solution” for analysis. Although TBNA is not widely used, it appears to improve the diagnostic yield when sampling from visible endobronchial, submucosal and peripheral lesions. Additionally the technique may detect mediastinal disease potentially allowing the operator to diagnose and stage a lung tumour in one procedure carried out under local anaesthetic.
TBNA may be used to sample Iymph nodes that lie immediately adjacent to the trachea and major bronchi. Care must be taken to perform TBNA prior to inspection of the distal airways and other sampling procedures as contamination with exfoliated malignant cells are a recognised cause of false-positive results. Studies have reported sensitivity rates of 43–83% and positive predictive values of 89–100% [66–71]. Use of a 19-gauge needle provides greater sensitivity than a 22-gauge needle but a combination of samples provides the best yields [65, 72]. Although the positive-predictive value is high (often 100%) the negative-predictive value is low and does not obviate the requirement for further surgical staging [62–72].
A potential limitation of mediastinal lymph node staging with TBNA is that the operator is blind. The technique may be combined with that of endobronchial ultrasound by the miniprobes [73, 74].
The numerous papers on TBNA performed in the last few years confirm the safety of the procedure. No cases of mortality related to TBNA have been described. The rare complications reported are: pneumothorax [75], pneumomediastinum [76], haemomediastinum [77], bacteraemia [78] and pericarditis [79]. None of these complications determined clinical major consequences.
One of the major complications of TBNA is the possible severe damage to the working channel of the scope [64]. This is more frequent with nonretractable needle (that should no longer be available in the market) and can be avoided if the operator takes care in introducing and extracting the needle from the bronchoscope with the tip completely retracted in the sheath.
Endo-oesophageal ultrasound with fine needle aspiration
Endo-oesophageal ultrasound with fine needle aspiration (EUS-FNA) is a relatively new method, first described in 1991 [80]. Since then, several studies have been published and it has been demonstrated that generally all lesions outlined by EUS may be punctured, and even small lesions down to the size of 5 mm may be diagnosed [81].
EUS-FNA is performed with the aid of oesophagoscopy and a biopsy needle is passed through the working channel of the endoscope, through the oesophageal wall and guided ultrasonographically toward the lesion of interest in the mediastinum. The procedure is performed under local anaesthetic and conscious sedation. This method gives an excellent overview of mediastinal structures, including a good access to the para-oesophageal space, the aortico-pulmonary window, the sub-carinal region and the region around the left atrium (level 4, 5 and 7) [82–84] and then EUS has the advantage of being noninvasive, safe and cost effective [85]. However, an area anterior to the air-filled trachea cannot be visualised.
The procedure can be per formed under i.v. sedation using midazolam or propofol [82]. The echoendoscope is initially introduced up to the level of celiac axis and gradually withdrawn upwards for a detailed mediastinal imaging. Since the ultrasound waves are emitted parallel to the long axis of the endoscope, the entire needle can be visualised approaching a target in the sector-shaped sound field. Pulse-wave Doppler ultrasonography imaging is performed, whenever vascular structures were supposed in the pathway of the needle or adjacent to it, to correct the target line if necessary [84]. The needle is advanced through the wall of the oesophagus and guided into the target lesion. The central stylet is removed, and a special 10-mL syringe attached to the hub of the needle to apply suction as the needle is moved back and forth within the mass. The suction is released slowly and the needle assembly removed out of the biopsy channel. One to two needle passes were made to obtain adequate tissue [86, 87].
Visual assessment of mediastinal lymph nodes by EUS gave for various observers sensitivities of 0.54–0.75, specificities of 0.71–0.98, positive-predictive values of 0.46–0.77, and negative-predictive values of 0.85–0.93, in a total of >1,000 patients [88]. These studies varied widely with regard to the number of examined mediastinal lymph node levels, visual criteria for malignancy and patient population characteristics. Compared with CT, the detection rate of malignant lymph nodes is higher with EUS, with fewer false positive results [89].
EUS can assess mediastinal lymph nodes at most levels, particularly at levels 4 left, 5, 7, 8 and 9, as well as metastasis in the left adrenal gland. Levels 1, 2, 3 and 4 right are not always assessable, because of interference by air in the larger airways [82, 84]. When enlarged, however, detection is easier [81, 88]. Properties of lymph nodes indicating possible malignancy are a hypoechoic core, sharp edges, round shape and a long axis diameter >10 mm [88]. Signs of benignancy are a hyperechoic core (fat), central calcification (old granulomatous disease), ill-defined edges, a long and narrow shape and a long axis diameter up to 1 cm [89, 90].
False negative results may have been introduced by the sometimes poor lymph node sampling during EUS-FNA (sampling only the most suspicious nodes). Many outcomes have been supported by clinical instead of surgical follow-up. Despite these drawbacks, the clinical impact of EUS-FNA is illustrated by a change in the management of NSCLC after EUS-FNA in 66% of the patients or cancellation of 68 and 49% of the scheduled mediastinoscopies and thoracotomies, respectively.
According to Hunerbein et al. [86], EUS-FNA made an unexpected diagnosis in 30% of the procedures. In two studies with decision–analysis models, EUS-FNA was shown to be less expensive compared with mediastinoscopy for the assessment of the entire mediastinum or for subcarinal lymph nodes only [82, 91, 92].
Barawi et al. [93] prospectively studied the incidence of complications associated with EUS-FNA. In 842 mediastinal EUS-FNA procedures, 1 infection, 2 haemorrhages and 1 inexplicable transient hypotension were reported. EUS-FNA is relatively contraindicated in patients with a Zenker’s diverticulum or bleeding tendency [94].
FNA of a cystic mediastinal lesion should be avoided, or when necessary be preceded by prophylactic antibiotics [95].
EBUS-TBNA
Lymph node staging is also the main indication for use of the new EBUS-TBNA scope. An ultrasound transducer integrated into a bronchoscope with a separate working channel would potentially increase the yield of TBNA by allowing direct visualisation of needle placement within the area of interest. So a special ultrasonic puncture bronchoscope by integrating a convex probe at the tip of the flexible bronchoscope was developed. With this bronchoscope direct TBNA under real-time convex probe endobronchial ultrasonography (EBUS-TBNA bronchoscopy) guidance is now possible.
EBUS-TBNA is performed by direct transducer contact with the wall of the trachea or bronchus. When a lesion is outlined, a 22-gauge full-length steel needle is introduced through the biopsy channel of the endoscope. Power Doppler examination may be performed before the biopsy to avoid unintended puncture of vessels. Under real-time ultrasonic guidance, the needle is placed in the lesion (fig. 4). Suction is applied with a syringe and the needle is moved back and forth inside the lesion [96].
Real-time ultrasonic guidance used to insert a needle into an identified lesion.
EBUS-TBNA has now been available for more than 5 years. A growing body of research supports its usefulness in airway assessment and procedure guidance, especially since PET is available [97–100].
EBUS-TBNA has access to all of the mediastinal lymph node stations accessible by mediastinoscopy as well as N1 nodes. The largest trial reported the results of the method in 502 patients [101]. 572 lymph nodes were punctured and 535 (94%) resulted in a diagnosis. Biopsies were taken from all reachable lymph node stations (2l, 2r, 3, 4r, 4l, 7, 10r, 10l, 11r and 11l. Mean±sd diameter of the nodes was 1.6±0.36 cm and the range was 0.8–3.2 cm. Sensitivity was 92%, specificity was 100% and the positive-predictive value was 93%. As in all other trials, no complications occurred.
A Danish–German group [102] additionally examined the accuracy of EBUS-TBNA in sampling nodes <1 cm in diameter. Among 100 patients, 119 lymph nodes with a size of 4–10 mm were detected and sampled. Malignancy was detected in 19 patients but missed in two others; all diagnoses were confirmed by surgical findings. The mean diameter of the punctured lymph nodes was 8.1 mm. The sensitivity of EBUS-TBNA for detecting malignancy was 92.3 %; the specificity was 100%; and the negative-predictive value was 96.3 %. Again, no complications occurred. They summarised that EBUS-TBNA can sample even small mediastinal nodes, therefore avoiding unnecessary surgical exploration in 1 out of 5 patients who have no CT evidence of mediastinal disease. Potentially operable patients with clinically nonmetastatic NSCLC may benefit from presurgical EBUS-TBNA biopsies and staging. A study comparing EBUS-TBNA, CT and PET for lymph node staging of lung cancer showed a high yield for EBUS-TBNA [103]. Altogether, 102 potentially operable patients with proven (n = 96) or radiologically suspected (n = 6) lung cancer were included in the study. CT, PET and EBUS-TBNA were performed prior to surgery for the evaluation of mediastinal and hilar lymph node metastasis. The sensitivities of CT, PET and EBUS-TBNA for the correct diagnosis of mediastinal and hilar lymph node staging were 76.9, 80.0 and 92.3%, respectively; the specificities were 55.3, 70.1 and 100%, respectively; and the diagnostic accuracies were 60.8, 72.5 and 98.0%, respectively. EBUS-TBNA was proven to have high sensitivity and specificity, compared with CT or PET, for mediastinal staging in patients with potentially resectable lung cancer.
Restaging of the mediastinum is another area of growing interest for the treatment strategy of lung cancer. In cases of advanced lymph node stage lung cancer, induction chemotherapy prior to surgical resection is an option. Mediastinoscopy is considered the gold standard for staging the mediastinum. However, re-mediastinoscopy can be technically difficult and is therefore not commonly performed. The ability to perform multiple, repeat biopsies using EBUS-TBNA allows restaging of the mediastinum after the introduction of chemotherapy.
A group of 124 consecutive patients with tissue-proven IIIA-N2 disease who were treated with induction chemotherapy underwent mediastinal restaging by EBUS-TBNA. The sensitivity, specificity, positive predictive value, negative predictive value, and diagnostic accuracy of EBUS-TBNA for mediastinal restaging following induction chemotherapy were 76, 100, 100, 20 and 77%, respectively. EBUS–TBNA is an accurate, minimally invasive test for mediastinal restaging of patients with NSCLC. However, because of the low negative-predictive value, tumour-negative findings should be confirmed by surgical staging [104].
EBUS-TBNA can be also used for the diagnosis of intrapulmonary nodules as well as mediastinal and hilar lymph nodes. The limitation is the reach of EBUS-TBNA, which depends on the size of the bronchus. Usually the EBUS-TBNA can be inserted as far as the lobar bronchus. Lung tumours located adjacent to the airway within reach of EBUS-TBNA can be diagnosed with EBUS-TBNA. Tornouy et al. [105] have reported their experience is this indication. In 60 patients, who have had a nondiagnostic bronchoscopy before, they were able to establish the defintive diagnosis in 77% without any complication.
Complications related to the procedure are similar to those of conventional TBNA, including bleeding from major vessels, pneumomediastinum, mediastinitis, pneumothorax, bronchospam and laryngospasm. Complications related to EBUS-TBNA have not been observed and to date there are no major complications reported in the literature. Although EBUS has enabled the bronchoscopist to see beyond the airway, one must be aware of the possible complications related to the procedure [106, 107].
Rapid on-site evaluation
Rapid on-site evaluation (ROSE) is comparable with the intra-operative frozen-section examination. The technique requires the cytopathologist and the pathology technician to process and interpret the stained wet film of the aspirate immediately and report the result to the bronchoscopist. For all needle techniques advantages are to resume. Several studies have shown that ROSE reduces the incidence of inadequate specimens, an important cause of nondiagnostic TBNA aspirates [108–111].
Davenport [112] studied the value of ROSE in 73 aspirates and compared the results with 134 specimens processed routinely. The aspirates were obtained from the mediastinal lymph nodes and the peripheral lung nodules. With ROSE, the proportion of aspirates showing malignant cells increased from 31% to 56%. The proportion of the inadequate negative specimens dropped from 56% to 18%. The negative aspirate with ROSE had a higher negative-predictive value than that of routinely prepared specimens. In a recent prospective study, Diette et al. [113] evaluated transbronchial needle aspirates with ROSE in 81 of 204 cases. The overall diagnostic yield was 81% when ROSE was used compared with a 50% yield when specimens were processed in the usual manner. Multivariate analysis showed that ROSE was an independent predictor of a positive aspirate for malignant cells with an odds ratio of 4.5. The mean number of needle attempts was slightly greater with ROSE. The concordance between the preliminary diagnosis made in the bronchoscopy suite and the final diagnosis was reached after subsequent review of material in the cytopathology laboratory was 87% indicating that the onsite evaluation of needle aspirate is fairly accurate but not perfect.
Although ROSE seems to improve the diagnostic yield of TBNA, its cost-effectiveness remains unclear. Successful use of ROSE requires services of an expert cytopathologist. Many pathologists do not favour ROSE because of the extra time and effort involved. The reimbursement from a third-party payer for these services is highly variable. Presently, the decision to use ROSE should be made on a case-by-case basis.
Footnotes