Low bacterial contamination of nebulizers in home treatment of cystic fibrosis patients

https://doi.org/10.1016/S0195-6701(97)90195-XGet rights and content

Abstract

Many reports have shown bacterial contamination of nebulizers used by patients with cystic fibrosis (CF) at home. At the Stockholm CF centre we recommend dismantling the equipment, washing, rinsing and drying it after each use, and disinfecting it once daily by boiling water or by 2% acetic acid followed by drying without rinsing. We studied whether patients comply with these recommendations and whether they are sufficient to prevent bacterial contamination. Nebulizers from 49 CF patients were investigated, 21 of whom are chronically colonized with Pseudomonas aeruginosa and one with Burkholderia cepacia. All patients were visited at home. Thirty-nine patients (79%) disinfected their equipment after the latest use in accordance with our recommendations. Thirty-eight pieces showed no, or only scanty, growth of micro-organisms belonging to the normal oropharyngeal flora. A moderate growth of α-streptococci was observed from a further five pieces. Four of these had not been cleaned after the latest inhalation occasion, and one was visibly dirty. Pseudomonads were observed from three pieces; two of these had been disinfected by boiling water and one by acetic acid, followed by rinsing in tap water. All three had been inadequately dried. Pseudomonads could not be cultured simultaneously from the sputum of these three patients. In conclusion, most patients comply with our cleaning and disinfection recommendations and these recommendations appear satisfactory in preventing bacterial contamination.

References (17)

There are more references available in the full text version of this article.

Cited by (44)

  • Nebulizer cleaning and disinfection practices in families with cystic fibrosis: The relationship between attitudes, practice and microbe colonization

    2019, Journal of Cystic Fibrosis
    Citation Excerpt :

    Surveys of CF patients show that most are aware of the importance of home nebulizer maintenance and the potential for germs to spread via aerosol [9,10]. However, patient adherence to recommended cleaning and disinfection practices is highly variable worldwide, ranging from 15% to 90% [9,11–14]. There are multiple reasons that likely contribute to difficulties following the CFF IP&C guidelines for home nebulizer care.

  • Microbial control and safety in inhalation devices

    2013, Inhaler Devices: Fundamentals, Design and Drug Delivery
  • Cystic fibrosis: Instrumental airway clearance techniques

    2012, Revue des Maladies Respiratoires
View all citing articles on Scopus
View full text